Configuration
R&S
®
FSW-K7
100User Manual 1173.9240.02 ─ 17
Trigger 2/3...................................................................................................................100
└ Output Type.................................................................................................. 100
└ Level................................................................................................... 100
└ Pulse Length.......................................................................................101
└ Send Trigger.......................................................................................101
Trigger 2/3
Defines the usage of the variable TRIGGER INPUT/OUTPUT connectors, where:
"Trigger 2": TRIGGER INPUT/OUTPUT connector on the front panel
"Trigger 3": TRIGGER 3 INPUT/ OUTPUT connector on the rear panel
(Trigger 1 is INPUT only.)
Note: Providing trigger signals as output is described in detail in the R&S FSW User
Manual.
"Input"
The signal at the connector is used as an external trigger source by
the R&S FSW. No further trigger parameters are available for the
connector.
"Output"
The R&S FSW sends a trigger signal to the output connector to be
used by connected devices.
Further trigger parameters are available for the connector.
Note: For offline AF or RF triggers, no output signal is provided.
Remote command:
OUTPut:TRIGger<port>:LEVel on page 253
OUTPut:TRIGger<port>:DIRection on page 253
Output Type ← Trigger 2/3
Type of signal to be sent to the output
"Device Trig-
gered"
(Default) Sends a trigger when the R&S FSW triggers.
"Trigger
Armed"
Sends a (high level) trigger when the R&S FSW is in "Ready for trig-
ger" state.
This state is indicated by a status bit in the STATus:OPERation reg-
ister (bit 5), as well as by a low level signal at the AUX port (pin 9).
For details see the description of the STATus:OPERation register in
the R&S FSW User Manual and the description of the AUX port in the
R&S FSW Getting Started manual.
"User Defined"
Sends a trigger when user selects "Send Trigger" button.
In this case, further parameters are available for the output signal.
Remote command:
OUTPut:TRIGger<port>:OTYPe on page 254
Level ← Output Type ← Trigger 2/3
Defines whether a constant high (1) or low (0) signal is sent to the output connector.
Remote command:
OUTPut:TRIGger<port>:LEVel on page 253
Trigger Configuration