EasyManua.ls Logo

Rohde & Schwarz FSW Series - Page 100

Rohde & Schwarz FSW Series
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration
R&S
®
FSW-K7
100User Manual 1173.9240.02 ─ 17
Trigger 2/3...................................................................................................................100
Output Type.................................................................................................. 100
Level................................................................................................... 100
Pulse Length.......................................................................................101
Send Trigger.......................................................................................101
Trigger 2/3
Defines the usage of the variable TRIGGER INPUT/OUTPUT connectors, where:
"Trigger 2": TRIGGER INPUT/OUTPUT connector on the front panel
"Trigger 3": TRIGGER 3 INPUT/ OUTPUT connector on the rear panel
(Trigger 1 is INPUT only.)
Note: Providing trigger signals as output is described in detail in the R&S FSW User
Manual.
"Input"
The signal at the connector is used as an external trigger source by
the R&S FSW. No further trigger parameters are available for the
connector.
"Output"
The R&S FSW sends a trigger signal to the output connector to be
used by connected devices.
Further trigger parameters are available for the connector.
Note: For offline AF or RF triggers, no output signal is provided.
Remote command:
OUTPut:TRIGger<port>:LEVel on page 253
OUTPut:TRIGger<port>:DIRection on page 253
Output Type ← Trigger 2/3
Type of signal to be sent to the output
"Device Trig-
gered"
(Default) Sends a trigger when the R&S FSW triggers.
"Trigger
Armed"
Sends a (high level) trigger when the R&S FSW is in "Ready for trig-
ger" state.
This state is indicated by a status bit in the STATus:OPERation reg-
ister (bit 5), as well as by a low level signal at the AUX port (pin 9).
For details see the description of the STATus:OPERation register in
the R&S FSW User Manual and the description of the AUX port in the
R&S FSW Getting Started manual.
"User Defined"
Sends a trigger when user selects "Send Trigger" button.
In this case, further parameters are available for the output signal.
Remote command:
OUTPut:TRIGger<port>:OTYPe on page 254
Level ← Output Type ← Trigger 2/3
Defines whether a constant high (1) or low (0) signal is sent to the output connector.
Remote command:
OUTPut:TRIGger<port>:LEVel on page 253
Trigger Configuration

Table of Contents

Other manuals for Rohde & Schwarz FSW Series

Related product manuals