Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
253User Manual 1173.9240.02 ─ 17
11.4.7.2 Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
TRIGGER INPUT/OUTPUT connectors. The tasks for manual operation are described
in "Trigger 2/3" on page 100.
OUTPut:TRIGger<port>:DIRection...................................................................................253
OUTPut:TRIGger<port>:LEVel........................................................................................ 253
OUTPut:TRIGger<port>:OTYPe...................................................................................... 254
OUTPut:TRIGger<port>:PULSe:IMMediate.......................................................................254
OUTPut:TRIGger<port>:PULSe:LENGth.......................................................................... 254
OUTPut:TRIGger<port>:DIRection <Direction>
This command selects the trigger direction.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Direction> INPut
Port works as an input.
OUTPut
Port works as an output.
*RST: INPut
Manual operation: See "Trigger 2/3" on page 100
OUTPut:TRIGger<port>:LEVel <Level>
This command defines the level of the signal generated at the trigger output.
This command works only if you have selected a user defined output with OUTPut:
TRIGger<port>:OTYPe.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Level> HIGH
TTL signal.
LOW
0 V
*RST: LOW
Manual operation: See "Trigger 2/3" on page 100
See "Level" on page 100
Configuring the Measurement