EasyManua.ls Logo

Rohde & Schwarz FSW Series - Page 182

Rohde & Schwarz FSW Series
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
182User Manual 1173.9240.02 ─ 17
INPut:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
after an overload condition occured and the protection mechanism intervened.The
error status bit (bit 3 in the STAT:QUES:POW status register) and the INPUT OVLD
message in the status bar are cleared.
(For details on the status register see the R&S FSW User Manual).
The command works only if the overload condition has been eliminated first.
Usage: Event
INPut:
CONNector <ConnType>
Determines whether the RF input data is taken from the RF input connector or the
optional Analog Baseband I connector. This command is only available if the Analog
Baseband interface (R&S FSW-B71) is installed and active for input.
For more information on the Analog Baseband Interface (R&S FSW-B71) see the
R&S FSW I/Q Analyzer and I/Q Input User Manual.
Parameters:
<ConnType> RF
RF input connector
AIQI
Analog Baseband I connector
*RST: RF
Example:
INP:CONN:AIQI
Selects the analog baseband input.
Usage: SCPI confirmed
Manual operation: See "Input Connector" on page 58
INPut:COUPling <CouplingType>
This command selects the coupling type of the RF input.
Parameters:
<CouplingType> AC
AC coupling
DC
DC coupling
*RST: AC
Example:
INP:COUP DC
Usage: SCPI confirmed
Manual operation: See "Input Coupling" on page 56