Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
182User Manual 1173.9240.02 ─ 17
INPut:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
after an overload condition occured and the protection mechanism intervened.The
error status bit (bit 3 in the STAT:QUES:POW status register) and the INPUT OVLD
message in the status bar are cleared.
(For details on the status register see the R&S FSW User Manual).
The command works only if the overload condition has been eliminated first.
Usage: Event
INPut:
CONNector <ConnType>
Determines whether the RF input data is taken from the RF input connector or the
optional Analog Baseband I connector. This command is only available if the Analog
Baseband interface (R&S FSW-B71) is installed and active for input.
For more information on the Analog Baseband Interface (R&S FSW-B71) see the
R&S FSW I/Q Analyzer and I/Q Input User Manual.
Parameters:
<ConnType> RF
RF input connector
AIQI
Analog Baseband I connector
*RST: RF
Example:
INP:CONN:AIQI
Selects the analog baseband input.
Usage: SCPI confirmed
Manual operation: See "Input Connector" on page 58
INPut:COUPling <CouplingType>
This command selects the coupling type of the RF input.
Parameters:
<CouplingType> AC
AC coupling
DC
DC coupling
*RST: AC
Example:
INP:COUP DC
Usage: SCPI confirmed
Manual operation: See "Input Coupling" on page 56
Configuring the Measurement