Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
195User Manual 1173.9240.02 ─ 17
Parameters:
<PortType> 2 | 3
*RST: 2
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:PORT 3
Manual operation: See "Mixer Type" on page 68
[SENSe:
]CORRection:CVL:SELect <FileName>
This command selects the conversion loss table with the specified file name. If
<file_name> is not available, a new conversion loss table is created.
This command is only available with option B21 (External Mixer) installed.
Parameters:
<FileName> '<File name>'
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Manual operation: See "New Table" on page 64
See "Edit Table" on page 65
See "File Name" on page 66
[SENSe:]CORRection:CVL:SNUMber <SerialNo>
This command defines the serial number of the mixer for which the conversion loss
table is to be used. This setting is checked against the current mixer setting before the
table can be assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see [SENSe:]CORRection:CVL:SELect on page 195).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<SerialNo> Serial number with a maximum of 16 characters
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:MIX '123.4567'
Manual operation: See "Mixer S/N" on page 67
Programming Example: Working with an External Mixer
This example demonstrates how to work with an external mixer in a remote environ-
ment. It is performed in the Spectrum application in the default layout configuration.
Note that without a real input signal and connected mixer, this measurement will not
return useful results.
//--------------Preparing the instrument -----------
//Reset the instrument
Configuring the Measurement