Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
199User Manual 1173.9240.02 ─ 17
Parameters:
<State> ON
Automatic definition
OFF
Manual definition according to INPut:IQ:FULLscale[:
LEVel] on page 199
*RST: ON
Example:
INP:IQ:FULL:AUTO OFF
Manual operation: See "Full Scale Level Mode / Value" on page 92
INPut:
IQ:FULLscale[:LEVel] <PeakVoltage>
This command defines the peak voltage at the Baseband Input connector if the full
scale level is set to manual mode (see INPut:IQ:FULLscale:AUTO on page 198).
Parameters:
<PeakVoltage> 0.25 V | 0.5 V | 1 V | 2 V
Peak voltage level at the connector.
For probes, the possible full scale values are adapted according
to the probe's attenuation and maximum allowed power.
*RST: 1V
Example:
INP:IQ:FULL 0.5V
Manual operation: See "Full Scale Level Mode / Value" on page 92
INPut:IQ:TYPE <DataType>
This command defines the format of the input signal.