Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
234User Manual 1173.9240.02 ─ 17
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence...........................................................234
CALCulate<n>:UNIT:POWer...........................................................................................234
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel............................................................ 234
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel:OFFSet................................................235
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
Usage: Event
CALCulate<n>:UNIT:POWer <Unit>
This command selects the unit of the y-axis.
The unit applies to all measurement windows.
Parameters:
<Unit> DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
*RST: dBm
Example:
CALC:UNIT:POW DBM
Sets the power unit to dBm.
Manual operation: See "Unit" on page 87
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel <ReferenceLevel>
This command defines the reference level.
With a reference level offset ≠ 0, the value range of the reference level is modified by
the offset.
Parameters:
<ReferenceLevel> The unit is variable.
Range: see datasheet
*RST: 0 dBm
Example:
DISP:TRAC:Y:RLEV -60dBm
Usage: SCPI confirmed
Manual operation: See "Reference Level" on page 86