Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
244User Manual 1173.9240.02 ─ 17
Parameters:
<SweepCount>
When you set a sweep count of 0 or 1, the R&S FSW performs
one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range: 0 to 200000
*RST: 0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Usage: SCPI confirmed
Manual operation: See "Sweep/Average Count" on page 106
[SENSe:
]SWEep:POINts <SweepPoints>
This command defines the number of measurement points analyzed during a sweep.
Parameters:
<SweepPoints> Range: 101 to 100001
*RST: 1001
Example:
SWE:POIN 251
Usage: SCPI confirmed
Manual operation: See "Sweep Points" on page 106
11.4.7 Triggering
The following remote commands are required to configure a triggered measurement in
a remote environment. More details are described for manual operation in chapter 5.5,
"Trigger Configuration", on page 94.
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the Triggering Conditions...................................................................245
Configuring the Trigger Output..............................................................................253
Configuring the Measurement