Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
244User Manual 1173.9240.02 ─ 17
Parameters:
<SweepCount>
When you set a sweep count of 0 or 1, the R&S FSW performs
one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range: 0 to 200000
*RST: 0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Usage: SCPI confirmed
Manual operation: See "Sweep/Average Count" on page 106
[SENSe:
]SWEep:POINts <SweepPoints>
This command defines the number of measurement points analyzed during a sweep.
Parameters:
<SweepPoints> Range: 101 to 100001
*RST: 1001
Example:
SWE:POIN 251
Usage: SCPI confirmed
Manual operation: See "Sweep Points" on page 106
11.4.7 Triggering
The following remote commands are required to configure a triggered measurement in
a remote environment. More details are described for manual operation in chapter 5.5,
"Trigger Configuration", on page 94.
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
● Configuring the Triggering Conditions...................................................................245
● Configuring the Trigger Output..............................................................................253
Configuring the Measurement