Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
263User Manual 1173.9240.02 ─ 17
Manual operation: See "Deemphasis" on page 116
[SENSe:]FILTer<n>:DEMPhasis[:STATe] <State>
This command activates/deactivates the selected deemphasis for the specified evalua-
tion.
For details about deemphasis refer to "Deemphasis" on page 116.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
FILT:DEMP ON
Activates the selected deemphasis.
Manual operation: See "Deemphasis" on page 116
[SENSe:]FILTer<n>:HPASs:FREQuency[:ABSolute] <FilterType>
This command selects the high pass filter type for the specified evaluation.
For details on the high pass filters refer to "High Pass" on page 114.
Parameters:
<FilterType> 20 Hz | 50 Hz | 300 Hz
*RST: 300Hz
Default unit: Hz
Example:
FILT:HPAS:FREQ 300Hz
Selects the high pass filter for the demodulation bandwidth
range from 800 Hz to 8 MHz.
Manual operation: See "High Pass" on page 114
[SENSe:]FILTer<n>:HPASs:FREQuency:MANual <Frequency>
This command selects the cutoff frequency of the high pass filter for the specified eval-
uation.
For details on the high pass filters refer to "High Pass" on page 114.
Parameters:
<Frequency> numeric value
Range: 0 to 3 MHz
*RST: 15kHz
Example:
FILT:HPAS:FREQ:MAN 3MHz
The AF results are restricted to frequencies lower than 3 MHz.
Manual operation: See "High Pass" on page 114