Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
264User Manual 1173.9240.02 ─ 17
[SENSe:]FILTer<n>:HPASs[:STATe] <State>
This command activates/deactivates the selected high pass filter for the specified eval-
uation.
For details on the high pass filter refer to "High Pass" on page 114.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
FILT:HPAS ON
Activates the selected high pass filter.
Manual operation: See "High Pass" on page 114
[SENSe:]FILTer<n>:LPASs:FREQuency[:ABSolute] <FilterType>
This command selects the absolute low pass filter type for the specified evaluation
For details on the low pass filter refer to "Low Pass" on page 114.
Parameters:
<FilterType> 3kHz | 15kHz | 150kHz
*RST: 15kHz
Example:
FILT:LPAS:FREQ 150kHz
Selects the low pass filter for the demodulation bandwidth range
from 400 kHz to 16 MHz.
Manual operation: See "Low Pass" on page 114
[SENSe:]FILTer<n>:LPASs:FREQuency:MANual <Frequency>
This command selects the cutoff frequency of the low pass filter for the specified evalu-
ation.
For details on the low pass filter refer to "Low Pass" on page 114.
Parameters:
<Frequency> numeric value
Range: 0 to 3 MHz
*RST: 15kHz
Example:
FILT:LPAS:FREQ:MAN 150kHz
The AF results are restricted to frequencies lower than 150 kHz.
Manual operation: See "Low Pass" on page 114
[SENSe:]FILTer<n>:LPASs:FREQuency:RELative <FilterType>
This command selects the relative low pass filter type for the specified evaluation
For details on the low pass filter refer to "Low Pass" on page 114.
Configuring the Measurement