Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
294User Manual 1173.9240.02 ─ 17
11.7.3 Retrieving Result Summary Values
The result summary contains measurement values that are calculated from the trace
data.
For details see "Result Summary" on page 23.
CALCulate<n>:MARKer:FUNCtion:ADEMod:AFRequency[:RESult]?...................................294
CALCulate:MARKer:FUNCtion:ADEMod:AM[:RESult<t>]?................................................. 294
CALCulate:MARKer:FUNCtion:ADEMod:FM[:RESult<t>]?..................................................294
CALCulate:MARKer:FUNCtion:ADEMod:PM[:RESult<t>]?................................................. 294
CALCulate:MARKer:FUNCtion:ADEMod:CARRier[:RESult]?.............................................. 295
CALCulate:MARKer:FUNCtion:ADEMod:FERRor[:RESult<t>]?...........................................295
CALCulate<n>:MARKer:FUNCtion:ADEMod:SINad:RESult<t>?..........................................295
CALCulate<n>:MARKer:FUNCtion:ADEMod:THD:RESult<t>?............................................295
[SENSe:]ADEMod:FM:OFFSet?...................................................................................... 296
[SENSe:]ADEMod:PM:RPOint[:X]....................................................................................297
CALCulate<n>:MARKer:FUNCtion:ADEMod:AFRequency[:RESult]?
This command queries the modulation (audio) frequency for the demodulation method
in the selected window.
Parameters:
<ModFreq> Modulation frequency in Hz.
Usage: Query only
CALCulate:MARKer:FUNCtion:ADEMod:AM[:RESult<t>]? <MeasType>
CALCulate:MARKer:FUNCtion:ADEMod:FM[:RESult<t>]? <MeasType>
CALCulate:MARKer:FUNCtion:ADEMod:PM[:RESult<t>]? <MeasType>
This command queries the current value of the demodulated signal for the specified
trace (as displayed in the Result Summary in manual operation).
Note that all windows with the same evaluation method have the same traces, thus the
window is irrelevant.
Query parameters:
<MeasType> PPEak | MPEak | MIDDle | RMS
PPEak
Postive peak (+PK)
MPEak | NPEak
Negative peak (-PK)
MIDDle
Average of positive and negative peaks ±PK/2
RMS
Root mean square value