Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
301User Manual 1173.9240.02 ─ 17
CALCulate<n>:MARKer<m1>:LINK:TO:MARKer<m2> <State>
This command links normal marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, marker <m1> changes its hori-
zontal position to the same value.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK4:LINK:TO:MARK2 ON
Links marker 4 to marker 2.
Manual operation: See "Linking to Another Marker" on page 136
CALCulate<n>:MARKer<m>[:STATe] <State>
This command turns markers on and off. If the corresponding marker number is cur-
rently active as a deltamarker, it is turned into a normal marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation: See "Marker State" on page 136
See "Marker Type" on page 136
CALCulate<n>:MARKer<m>:TRACe <Trace>
This command selects the trace the marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Parameters:
<Trace> 1 to 6
Trace number the marker is assigned to.
Example:
CALC:MARK3:TRAC 2
Assigns marker 3 to trace 2.
Manual operation: See "Assigning the Marker to a Trace" on page 137
CALCulate<n>:MARKer<m>:X <Position>
This command moves a marker to a particular coordinate on the x-axis.
If necessary, the command activates the marker.
Analyzing Results