Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
305User Manual 1173.9240.02 ─ 17
Parameters:
<Trace> Trace number the marker is assigned to.
Example:
CALC:DELT2:TRAC 2
Positions delta marker 2 on trace 2.
CALCulate<n>:DELTamarker<m>:X <Position>
This command moves a delta marker to a particular coordinate on the x-axis.
If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
The position is relative to the reference marker.
To select an absolute position you have to change the delta
marker mode with CALCulate<n>:DELTamarker:MODE
on page 303.
A query returns the absolute position of the delta marker.
Range: The value range and unit depend on the measure-
ment and scale of the x-axis.
Example:
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
Manual operation: See "Marker Position (X-value)" on page 136
CALCulate<n>:DELTamarker<m>:X:RELative?
This command queries the relative position of a delta marker on the x-axis.
If necessary, the command activates the delta marker first.
Return values:
<Position> Position of the delta marker in relation to the reference marker or
the fixed reference.
Example:
CALC:DELT3:X:REL?
Outputs the frequency of delta marker 3 relative to marker 1 or
relative to the reference position.
Usage: Query only
CALCulate<n>:DELTamarker<m>:Y?
This command queries the relative position of a delta marker on the y-axis.
If necessary, the command activates the delta marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 278.
Analyzing Results