Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
315User Manual 1173.9240.02 ─ 17
The actual number of peaks that have been found may differ from the number of peaks
you have set to be found because of the peak excursion.
Return values:
<NumberOfPeaks>
Example:
CALC:MARK:FUNC:FPE:COUN?
Queries the number of peaks.
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks[:IMMediate] <Peaks>
This command initiates a peak search.
Parameters:
<Peaks> This parameter defines the number of peaks to find during the
search.
Note that the actual number of peaks found during the search
also depends on the peak excursion you have set with
CALCulate<n>:MARKer:PEXCursion.
Range: 1 to 200
Example:
CALC:MARK:PEXC 5
Defines a peak excursion of 5 dB, i.e. peaks must be at least 5
dB apart to be detected as a peak.
CALC:MARK:FUNC:FPE 10
Initiates a search for 10 peaks on the current trace.
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE <MaxNoPeaks>
This command defines the maximum number of peaks that the R&S FSW looks for
during a peak search.
Parameters:
<MaxNoPeaks> Maximum number of peaks to be determined.
Range: 1 to 200
*RST: 50
Example:
CALC:MARK:FUNC:FPE:LIST:SIZE 10
The marker peak list will contain a maximum of 10 peaks.
Manual operation: See "Maximum Number of Peaks" on page 146
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT <SortMode>
This command selects the order in which the results of a peak search are returned.
Analyzing Results