Example:
CALC:DELT:FUNC:FIX ON
Switches on the measurement with fixed reference value for all
delta markers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm.
Manual operation: See "Defining a Fixed Reference" on page 138
Marker Peak Lists
Useful commands for peak lists described elsewhere
CALCulate<n>:MARKer:PEXCursion on page 308
MMEMory:STORe:PEAK on page 317
chapter 11.8.1.3, "Marker Search Settings", on page 308
Remote commands exclusive to peak lists
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]........................ 314
CALCulate:MARKer:FUNCtion:FPEaks:COUNt?...............................................................314
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks[:IMMediate].............................................. 315
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE................................................ 315
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT...................................................... 315
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STAT.......................................................316
CALCulate:MARKer:FUNCtion:FPEeaks:X?..................................................................... 316
CALCulate:MARKer:FUNCtion:FPEeaks:Y?..................................................................... 316
MMEMory:STORe:LIST..................................................................................................317
MMEMory:STORe:PEAK................................................................................................ 317
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
<State>
This command turns labels for peaks found during a peak search on and off.
The labels correspond to the marker number in the marker peak list.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
CALC:MARK:FUNC:FPE:ANN:LAB:STAT OFF
Removes the peak labels from the diagram
Manual operation: See "Displaying Marker Numbers" on page 147
CALCulate:MARKer:FUNCtion:FPEaks:COUNt?
This command queries the number of peaks that have been found during a peak
search.