Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
331User Manual 1173.9240.02 ─ 17
CALCulate:LIMit<k>:TRACe<t>:CHECk <State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use CALCulate:LIMit<k>:FAIL?.
Note that this command replaces the two commands from previous signal and spec-
trum analyzers (which are still supported, however):
CALC:LIM:TRAC; see the description of commands for compatibility in the
R&S FSW User Manual
CALCulate:LIMit<k>:STATe on page 330
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM3:TRAC2:CHEC ON
Switches on the limit check for limit line 3 on trace 2.
Manual operation: See "Traces to be Checked" on page 149
11.8.2.3 Checking the Results of a Limit Check
CALCulate:LIMit:CLEar[:IMMediate].................................................................................331
CALCulate:LIMit<k>:FAIL?..............................................................................................331
CALCulate:LIMit:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on all limit lines in all measurement windows at the same time.
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
Usage: SCPI confirmed
CALCulate:LIMit<k>:FAIL?
This command queries the result of a limit check.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 278.
Return values:
<Result> 0
PASS
1
FAIL
Analyzing Results