EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 111

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Functions
NRP2
108User Manual 1173.9140.02 ─ 07
Table 4-5: Formulas for sum and difference of power
Function
Unit
W dBm dbμV
Sum (A+B)
BA
PP
mW1
lg10
BA
PP
Vμ1
lg20
ZPP
BA
Diff (A-B)
BA
PP
mW1
lg10
BA
PP
Vμ1
lg20
ZPP
BA
Table 4-6: Formulas for relative measurement
Function
Unit
1 dB Δ%
Single
R
P
P
R
P
P
lg10
1100
R
P
P
Sum
Diff
R
BA
P
PP
mW1
lg10
BA
PP
1100
R
BA
P
PP
Ratio
R
Y
X
B
A
P
P
P
P
R
Y
X
B
A
P
P
P
P
lg10
1100
R
Y
X
B
A
P
P
P
P
Table 4-7: Legend for formula tables
Variable Meaning
P
P
A
P
B
P
R
(P
X
/ P
Y
)
R
Z
Power
Power in channel A (incident power)
Power in channel B (reflected power)
Reference (power)
Reference (power ratio)
Characteristic impedance, i.e. 50Ω
Unit
The "Unit" drop-down list is used to set the result unit. Units that are provided for the
selected measurement function:
"W"
Single, Diff, Sum
"dBm | dBμV"
Single, Diff, Sum
"Δ% | dB | 1"
Ratio
Configuring measurements