The "Test..." function offers tests for the internal circuitry of the sensor (selftest 1) as
well as simple tests with applied RF (option R&S NRP-B1/-B7 required).
4.7.6.1 Internal sensor selftest
Open the "Sensortest" dialog box in the "System""Test..." menu.
Fig. 4-93: Sensortest dialog box / page 1
The first page of the "Sensortest" dialog box covers the settings for performing an
internal selftest of the sensor without using the reference signal.
A | B | C | D
Tabs for selecting the sensor.
Note: In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix 1 for channel A, channel 2 for B, and so on.
Selftest internal (with no signal)
Starts a selftest of the sensor. A signal must not be applied.
The instrument indicates the corresponding states to internal self-test directly in the
menu:
"...not tested" (prior to testing)
"...running" (in progress)
"...successful" (outcome)
"...failed" (outcome)
Test Report…
Opens the "Sensor Selftest Report"
window displaying the results of the internal selft-
est.
The status bar of the report window shows type and serial number of the tested sen-
sor, and the overall test result. The results of the individual tests are indicated below.