EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 166

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Functions
NRP2
163User Manual 1173.9140.02 ─ 07
Fig. 4-94: Sensor Selftest Report dialog box
4.7.6.2 Sensor test with applied RF
Testing the sensor with the RF reference signal requires option R&S NRP-B1/-B7,
Sensor Check Source.
Open the "Sensortest" dialog box in the "System""Test..." menu and select the
second page.
Fig. 4-95: Sensortest dialog box / page 2
This page of the "Sensortest" dialog box provides testing the sensors with the refer-
ence signal supplied by the R&S NRP-B1/-B7 test generator option.
In the case of sensors with a lead-in attenuator (pad) (R&S NRP-Z22, R&S NRP-Z23
and R&S NRP-Z24), the test measurement can be made with or without an attenuator.
Since the power sensors of the R&S NRP-Z series are absolutely calibrated, a regular
calibration measurement after switching on the instrument is not necessary and also
not planned.
The outcome of the tests is not considered during subsequent measurements.
Test at Power Ref
Starts a test measurement for a sensor. The sensor must be connected to the sensor
check source.
If an attenuator is present, it must first be removed.
System settings
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com

Table of Contents

Related product manuals