Remote Control - Commands
NRP2
201User Manual 1173.9140.02 ─ 07
Command Preset and *RST value Remark
OUTP:TTL:ACT LOW
OUTPut:TTL:LVOLtage
is
applied to the TTL output if a limit
is violated.
OUTP:TTL:
FAIL HIGH
OUTPut:TTL:HVOLtage is
applied to the TTL output if a limit
is violated.
OUTP:TTL:
FEED "CALC1:LIM"
The TTL output is fed by the limit
monitoring function of calculate
block 1.
OUTP:TTL:HVOLT 3.3 V
High voltage at TTL output
OUTP:TTL:LVOLT 0.0 V
Low voltage at TTL output
SENS<[1]..4>:AC:RANG 1 | 2 | 3
Depending on sensor.
SENS<[1]..4>:AC:RANG:AUT ON | OFF
Depending on sensor.
SENS<[1]..4>:AC:RANG:CLEV <NRf>
Depending on sensor.
SENS<[1]..4>:APER <NRf>
Depending on sensor.
SENS<[1]..4>:AUXiliary NONE | MINM
Depending on sensor.
SENS<[1]..4>:AVER ON | OFF
Depending on sensor.
SENS<[1]..4>:AVER:COUN <NR1>
Depending on sensor.
SENS<[1]..4>:AVER:COUN:AUTO ON | OFF
Depending on sensor.
SENS<[1]..4>:AVER:COUN:AUTO:
MTIMe
<NRf>
Depending on sensor.
SENS<[1]..4>:AVER:COUN:AUTO:
RES
<NRf>
Depending on sensor.
SENS<[1]..4>:AVER:COUN:AUTO:
SLOT
<NR1>
Depending on sensor.
SENS<[1]..4>:AVER:COUN:AUTO:
NSR
<NRf>
Depending on sensor.
SENS<[1]..4>:AVER:COUN:AUTO:
TYPE
RES | NSR
Depending on sensor.
SENS<[1]..4>:AVER:TCON MOV
REP
PRES and SYSTem:
PRESet
.
*RST
SENS<[1]..4>:BAND:VID
SENS<[1]..4>:BWID:VID
<string>
Depending on sensor.
SENS<[1]..4>:BUFF:SIZe <NR1>
Depending on sensor.
SENS<[1]..4>:BUFF:STAT ON | OFF
Depending on sensor.
SENS<[1]..4>:BURS:CHOP ON | OFF
Depending on sensor.
SENS<[1]..4>:BURS:DTOL <NRf>
Depending on sensor.
SENS<[1]..4>:CHOP ON | OFF
Depending on sensor.
SENS<[1]..4>:CORR:DCYC <NRf>
Depending on sensor.
Common Commands to IEEE 488.2