Remote Control - Commands
NRP2
265User Manual 1173.9140.02 ─ 07
However, SENS<[1]..4>:AVER:COUN:
AUTO:RES does not affect the DISPlay
command. The parameters of the two commands are different but have the same
meaning.
SENS<[1]..4>:AVER:COUN:AUTO:RES
1 2 3 4
DISP<[1]..4>:RES
1 0.1 0.01 0.001
Parameters:
<<resolution>> 1 | 2 | 3 | 4
Range: depending on sensor
*
RST: depending on sensor
[SENSe<[1]..4>]:AVERage:
COUNt:AUTO:SLOT <slot>
Sets a timeslot whose measured value is used to automatically determine the filter
length.
Parameters:
<slot> int_value
Range: depending on sensor
*
RST: depending on sensor
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:NSRatio <nsratio>
Indicates the maximum noise ratio in the measurement result. Achieving the same
result by filter length setting is attempted only if
SENS<[1]..4>:AVER:
COUN:AUTO ON and
SENS<[1]..4>:AVER:COUN:AUTO:TYPE NSR are set.
Unit: DB | PCT
Parameters:
<nsratio> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
DB
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:TYPE <mode>
Selects a method by which the automatic filter length switchover can operate: