Remote Control - Commands
NRP2
266User Manual 1173.9140.02 ─ 07
Parameters:
<mode> RESolution
takes setting SENS<[1]..4>:
AVER:
COUN:AUTO:RES into
account.
NSR
takes setting SENS<[1]..4>:AVER:COUN:AUTO:NSR into
account.
*RST: depending on sensor
[SENSe<[1]..4>]:AVERage:RESet
Initializes the digital filter by deleting the stored measured values.
Usage: Event
[SENSe<[1]..4>]:AVERage[:STATe] <state>
Activates or deactivates the filter function of a sensor. If the filter is switched on, the
individual measured values set with SENSe<[1]..4>]:AVERage:
COUNt are aver-
aged. This reduces the impact of noise and helps to achieve measured values that are
more stable.
Parameters:
<state> ON | OFF
*RST: depending on sensor
[SENSe<[1]..4>]:AVERage:TCONtrol <mode>
As soon as a new single value is determined, the filter window is advanced by one
value so that the new value is taken into account by the filter and the oldest value is
forgotten. [SENSe<[1]..4>]:AVERage:
TCONtrol (terminal control) then deter-
mines whether a new result is calculated immediately after a new measured value is
available (MOVing) or only after an entire range of new values is available for the filter
(REPeat).
Note:
In the MOVing mode, data from a point of time before the start of the current measure-
ment can influence the measurement result if the data is still in the filter.
Parameters:
<mode> MOVing | REPeat
*RST: REPeat; (PRES) and SYST:PRES: MOVing
[SENSe<[1]..4>]:BANDwidth:VIDeo <bandwidth>
[SENSe<[1]..4>]:BWIDth:VIDeo <bandwidth>
Sets the video bandwidth for the Trace mode.
SENSe (Sensor Commands)