Remote Control - Commands
NRP2
267User Manual 1173.9140.02 ─ 07
Parameters:
<bandwidth> string
Range: "FULL", "5 MHz", "1.5 MHz", "300 kHz"
*
RST: depending on sensor
[SENSe<[1]..4>]:BANDwidth:VIDeo:LIST?
[SENSe<[1]..4>]:BWIDth:VIDeo:LIST?
Provides a list of strings with permissible settings for
the
[SENSe<[1]..4>]:BWIDth:VIDeo command. Lists are not affected by the
*RST
command.
Usage: Query only
[SENSe<[1]..4>]:CORRection:OFFSet <offset_value>
A fixed offset value can be defined for multiplying (logarithmically adding) the mea-
sured value of a sensor. If the parameters are specified without a unit, the unit set by
UNIT:POW:RAT
will be used.
Unit: DB | PCT
Parameters:
<offset_value> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
DB
[SENSe<[1]..4>]:CORRection:OFFSet:STATe <state>
Switches the offset correction on or off.
Parameters:
<state> ON | OFF
*
RST: depending on sensor
[SENSe<[1]..4>]:CORRection:DCYCle[:INPut][:MAGNitude] <magnitude>
Informs the R&S NRP2 about the duty cycle of the power to be measured. The duty
cycle correction is only applied in the ContAv mode where measurements are per-
formed continuously without taking the timing of the signal into account.
Parameters:
<magnitude> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
PCT
SENSe (Sensor Commands)