270User Manual 1173.9140.02 ─ 07
To increase measurement accuracy, the sensors measure each signal twice in the
Burst, Timeslot and Trace measurement modes (exception: in the Trace mode, with
SENS:TRAC:REAL ON). As long as the trigger system is set to internal or external trig-
gering, this is usually of no importance to the user. If TRIG:SOUR
HOLD or BUS is
used, however, then each measurement must be triggered twice (with *TRG or
TRIG:IMM).
Error messages:
24 "Sensor mode not supported": A sensor does not support a measurement mode.
28 "Sensor not idle": The sensor is not in the IDLE state.
-151 "Invalid string data": An invalid parameter was transmitted for <string>.