Remote Control - Commands
NRP2
327User Manual 1173.9140.02 ─ 07
SYSTem:STANdard:CATalog?
Returns a list of stored standards containing a series of strings separated by commas.
The strings can be used as parameters for theSYSTem:STANdard:
PRESet
<string> command.
Usage: Query only
SYSTem:STANdard:PRESet <string>
The R&S NRP2 can be easily configured for measuring standardized signals (e.g.
mobile radio standards). For this purpose, the standards listed in the annex (chapter A.
3, "Standard Configurations", on page 368) are stored in the R&S NRP2. The
SYST:STAN:
PRES command sets the listed parameters and, if desired, also config-
ures front-panel display windows (if SYST:STAN:PWS ON).
Parameters:
<string> The available standards and the associated strings can be quer-
ied with the SYSTem:
STANdard:CATalog command.
*RST: <empty_string>
SYSTem:STANdard:PWSettings <state>
If SYST:STAN:PRES <string> is used, SYST:STAN:PWS (Preserve Window Set-
tings) determines whether the window system is also configured with the calculate
blocks (OFF) or whether exclusively sensor parameters are set (ON).
Parameters:
<state> ON | OFF
*RST: OFF
SYSTem:STANdard:TRIGger:SOURce <source>
If SYST:STAN:PRES <string> is used, this command determines whether the sen-
sors are set to internal (INT) or external (EXT) triggering.
Parameters:
<source> INT | EXT
*RST: INT
SYSTem:TIME <hour>,<minute>,<second>
Sets the time. Since the R&S NRP2 has no battery-backed clock, the time has to be
reset if the clock was not set after the last power-on of the R&S NRP2. (see also
SYSTem:
DATE
on page 322).
The time is set in the format (h)h,(m)m,(s)s. This setting is not changed by *RST.
SYSTem