Remote Control - Commands
NRP2
322User Manual 1173.9140.02 ─ 07
Parameters:
<sm1...4> int_value
Range: 0...255
*
RST: This setting is not changed by *RST.
Usage: Event
SYSTem:DATE
<year>,<month>,<day>
Sets the date. Since the R&S NRP2 has no battery-backed clock, the date has to be
reset if the clock was not set after the last power-on of the R&S NRP2.
The date is set in the format yyyy,(m)m,(d)d. This setting is not changed by *RST.
Table 6-33: Parameter range
Parameter Value range
<year> 2000 ... 2100
<month> 1 (January) ... 12 (December)
<day> 1 ... 31
SYSTem:IDN <string>
Determines the user defined return value for the *IDN query. The maximum string
length is 128 characters.
SYSTem:IDN:AUTO <state>
Activates/deactivates return of the user-defined string for the *IDN query.
SYSTem:INFO? <string>
Returns information about the system. <string_value> is used to query a specific
information item. If called without parameters, the command returns all available infor-
mation in the form of a list of strings separated by commas. <string_value> can
have the following values. No difference is made between upper-case and lower-case
letters:
<string_value> Meaning
"Manufacturer"
Manufacturer of device
"Type"
Type (R&S NRP2)
"Stock Number"
Order No. of R&S NRP2
"Serial"
Serial number
"HWVersion"
Hardware version
"SW Build"
Firmware version