Remote Control - Commands
NRP2
331User Manual 1173.9140.02 ─ 07
Example: (synchronized measurement with 2 heads):
*RST
TRIG:ALL:SLOP POS
TRIG:ALL:
SOUR EXT
TRIG:ALL:SYNC ON
INIT:ALL
FETCh1?; FETCh2?
TRIGger<[1]..4>:ATRigger[:STATe] <state>
TRIGger:ALL:ATRigger[:STATe] <state>
When TRIG:
ATR
is set to ON, the WAIT_FOR_TRG state is automatically exited when
no trigger event occurs within a period that corresponds to the reciprocal of the display
update rate.
Parameters:
<state> ON | OFF
*RST: depending on sensor
TRIGger<[1]..4>:COUNt <value>
TRIGger:ALL:COUNt <value>
Sets the number of measurement cycles to be performed when the measurement is
started with INIT.
Parameters:
<value> int_value
Range: depending on sensor
*
RST: depending on sensor
Default unit:
1
TRIGger<[1]..4>:DELay <delay>
TRIGger:ALL:DELay <delay>
Defines the delay between the trigger event and the beginning of the actual measure-
ment (integration).
Parameters:
<delay> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit:
S