Remote Control - Commands
NRP2
333User Manual 1173.9140.02 ─ 07
TRIGger<[1]..4>:HOLDoff <holdoff>
TRIGger:ALL:HOLDoff <holdoff>
Defines a period after a trigger event within which all further trigger events are ignored.
Parameters:
<holdoff> float_value
Range: depending on sensor
*
RST: depending on sensor
Default unit:
S
TRIGger<[1]..4>:HYSTeresis <value>
TRIGger:ALL:HYSTeresis <value>
This command is used to specify how far the signal level has to drop below the trigger
level before a new signal edge can be detected as a trigger event. Thus, this command
can be used to eliminate the effects of noise in the signal on the edge detector of the
trigger system.
Unit: DB | PCT
Parameters:
<value> float_value
Range: depending on sensor
*
RST: depending on sensor
Default unit:
DB
TRIGger<[1]..4>[:IMMediate]
TRIGger:ALL[:IMMediate]
Performs triggering and makes sure that the sensor directly changes from the
WAIT_FOR_TRG state to the MEASURING state irrespective of the selected trigger
source (TRIG:SOUR
). A trigger delay set with TRIG:DEL is ignored but not the auto-
matic delay determined when TRIG:DEL:AUTO:ON is set. When the trigger source is
HOLD, a measurement can only be started with TRIG.
Error messages:
-211 "Trigger ignored": The sensor is not in the WAIT_FOR_TRG state.
Usage: Event
TRIGger<[1]..4>:LEVel <level>
TRIGger:ALL:LEVel
<level>
Determines the power a trigger signal must exceed (TRIG:SLOP POS) or fall short of
(TRIG:SLOP NEG) before a trigger event is detected. This setting is only used for the
trigger signal source (TRIG:SOUR INTernal).
Unit: DBM | W
TRIGger