EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 337

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands
NRP2
334User Manual 1173.9140.02 ─ 07
Parameters:
<level> float_value
Range: depending on sensor
*
RST: depending on sensor
Default unit:
DBM
TRIGger<[1]..4>:MASTer[:STATe] <state>
Sets the selected sensor in trigger master state. If the trigger master detects an(inter-
nal) trigger event, it generates an external (hardware) trigger signal, which allows the
synchronous triggering of other sensors. The trigger master must be not be set to
external triggering, however, all other sensors have too be set to external triggering
(TRIG:SOUR EXT
). The trigger signal generated by the trigger master is routed via
the connecting cable to the R&S NRP2 and from there distributed to the other sensors
and the trigger output. Only one sensor can be trigger master at a time.
See also TRIG:SYNC
Parameters:
<state> ON | OFF
*RST: depending on sensor
TRIGger<[1]..4>:SLOPe <state>
TRIGger:ALL:SLOPe
<state>
This command determines whether the rising (POSitive) or the falling (NEGative)
edge of the signal is used for triggering.
Parameters:
<state> POSitive | NEGative
*RST: depending on sensor
TRIGger<[1]..4>:SOURce <state>
TRIGger:ALL:SOURce <state>
Sets the trigger signal source for the WAIT_FOR_TRG state.
TRIGger
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com

Table of Contents

Related product manuals