Error Messages and Troubleshooting
NRP2
341User Manual 1173.9140.02 ─ 07
Error code Error text returned upon queue query
Explanations
-113 "Undefined header"
The header is not defined for the device.
-114 "Header suffix out of range"
The header contains an illegal numeric suffix.
-120 "Numeric data error"
The command contains a faulty numeric parameter.
-121 "Invalid character in number"
A numeric contains an invalid character.
-123 "Exponent too large"
The magnitude of the exponent is larger than 32000.
-124 "Too many digits"
The numeric contains too many digits.
-127 "Invalid numeric data"
-128 "Numeric data not allowed"
The command contains a numeric data element in a position where it is not accepted.
-131 "Invalid suffix"
The suffix is invalid for this device. The suffix is the index that indexes the sensor with the
SENSe commands (1-4), the CALCulate block with the high-level measurement commands
(1-4) and the window with the DISPlay commands (1-4).
-134 "Suffix too long"
The suffix contains more than 12 characters.
-138 "Suffix not allowed"
A suffix is not allowed for this command or at this position in the command.
-141 "Invalid character data"
Either the character data element contains an invalid character or the particular element
received is not valid for this command.
-144 "Character data too long"
The character data element contains too many characters.
-148 "Character data not allowed"
The character data element used is not allowed for this command or at this position of the
command.
-151 "Invalid string data"
The command contains invalid string data.
-158 "String data not allowed"
The command contains a valid string data element at a position where it is not allowed.
-161 "Invalid block data"
The command contains invalid block data.