Instrument Functions
NRP2
77User Manual 1173.9140.02 ─ 07
Fig. 4-14: Offset dialog box
The Offset menu shows the calculated offset values. "Table Offset" is the offset calcu-
lated from the selected table and the current frequency. Its value does not depend on
whether table offsets are active or not.
"Sensor Offset"
shows the effective offset value.
This value takes into account the settings for "Global" and "Table" offset. However, it
neglects any additional offsets, that arise from the S-Parameter setting, as this infor-
mation is stored in the sensor.
A | B | C | D
Tabs for selecting the sensor.
Note: In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:..., i.e. choose suffix 1 for channel A, channel 2 for B, and so on.
Global (value)
Enters the global dB offset (dB offset).
Positive values correspond to external losses, negative values to external gains.
Remote command:
[SENSe<[1]..4>]:CORRection:
OFFSet on page 267
Global (state)
Activates/deactivates the global offset correction.
Remote command:
[SENSe<[1]..4>]:CORRection:OFFSet:STATe on page 267
Table (name)
Opens an item list containing correction tables.
The instrument provides 10 tables with up to 80 frequency / offset pairs each. The
default labels are "Table 1"
to "Table 10"; however, the names can be changed using
the table editor.
Remote command:
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect] on page 268