Command Reference
R&S
®
ZNB/ZNBT
1161User Manual 1173.9163.02 ─ 53
If the stop frequency entered is smaller than the current start frequency, the start fre-
quency is set to the stop frequency minus the minimum frequency span.
8.3.14.11 [SENSe:]HARMonic...
Implements functions related to harmonic grids for time domain transformation.
[SENSe<Ch>:]HARMonic?............................................................................................1161
[SENSe<Ch>:]HARMonic:AUTO....................................................................................1161
[SENSe<Ch>:]HARMonic:DLENgth:DATA.......................................................................1161
[SENSe<Ch>:]HARMonic:ELENgth:DATA.......................................................................1162
[SENSe<Ch>:]HARMonic:MLENgth:DATA...................................................................... 1162
[SENSe<Ch>:]HARMonic:PERMittivity:DATA.................................................................. 1162
[SENSe<Ch>:]HARMonic:RTIMe:DATA.......................................................................... 1163
[SENSe<Ch>:]HARMonic:RTIMe:THReshold.................................................................. 1163
[SENSe<Ch>:]HARMonic:VELocity:DATA....................................................................... 1163
[SENSe<Ch>:]HARMonic?
Queries whether the current frequency grid is harmonic.
Suffix:
<Ch>
.
Channel number
Return values:
<arg0>
Usage: Query only
Manual operation: See "Is the Current Grid Harmonic?" on page 315
[SENSe<Ch>:]HARMonic:AUTO <arg0>
Turns the "Automatic Harmonic Grid" function ON or OFF.
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0> *RST: ON
Options: R&S ZNB/ZNBT-K2
Manual operation: See "Automatic Harmonic Grid" on page 315
[SENSe<Ch>:]HARMonic:DLENgth:DATA <DUTLength>
Sets/gets the expected maximum time delay through the DUT.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference