Command Reference
R&S
®
ZNB/ZNBT
1238User Manual 1173.9163.02 ─ 53
SOURce<Ch>:POWer<PhyPt>:CORRection:PPOWer <RecCalPower>
Defines the source power which the R&S ZNB/ZNBT uses to perform the first calibra-
tion sweep of the source power calibration ("Reference Receiver Cal Power"). The
power value is ignored if the R&S ZNB/ZNBT is set to use the port power result
(SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect CPOW).
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<RecCalPower> Reference receiver cal power
Range: Depending on source power range of the analyzer
and the power range of the external power meter.
*RST: 0 dBm
Default unit: dBm
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation: See "Reference Receiver Cal Power" on page 495
SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect <CalPower>
Qualifies how to define the source power which the R&S ZNB/ZNBT uses to perform
the first calibration sweep of the source power calibration ("Reference Receiver Cal
Power").
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<CalPower> CPOWer | PPOWer
PPOWer use cal power value defined via SOURce<Ch>:
POWer<PhyPt>:CORRection:PPOWer
CPOWer use port power value
*RST: PPOWer
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation: See "Reference Receiver Cal Power" on page 495