402User Manual 1173.9163.02 ─ 53
Tip: Overlapping limit line and ripple limit line segments are not displayed when a
point-based x-axis is active; see Chapter 5.4.1.1, "Rules for Limit Line Definition",
on page 135.
Remote command:
[SENSe<Ch>:]FREQuency:SEGMent:AXIS
6.10.2.2 Define Segments Dialog
The "Define Segments" dialog defines all channel settings for a Segmented frequency
sweep and imports or exports segmented sweep settings.
Access: CHANNEL – [SWEEP] > "Sweep Type" > "Define Segments..."
The dialog contains a table to edit the individual segments of the sweep range. Use the
Displayed Columns Dialog to select the columns to be displayed and edited.
Sweep segments are allowed to overlap.
In contrast to standard frequency sweeps, the start and stop frequency in a sweep
segment must not be different.
So with a segmented sweep you can measure n
1
points at frequency f
1
(in segment
1), n
2
points at frequency f
2
(in segment 2) etc.