GUI Reference
R&S
®
ZNB/ZNBT
421User Manual 1173.9163.02 ─ 53
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Characterization
Displays all characterizations that are stored in the active cal unit. The "Factory" char-
acterization is available for all calibration units; it ensures an accurate calibration for all
standard applications. To account for modifications of the cal unit such as the connec-
tion of additional adapters, you can generate modified sets of characterization data
using the cal unit characterization wizard. See Chapter 6.11.2.3, "Characterize Cal Unit
Dialog", on page 486. By default, the R&S ZNB/ZNBT uses the last generated cal unit
characterization.
Tip: If the characterization wizard is password-protected, the "Characterization" button
is unavailable. Use this functionality to prevent inadvertent activation of inappropriate
characterizations. See "Authentication" on page 487.
See also Chapter 5.5.5.3, "Characterization of Calibration Units", on page 171.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Query further cal unit properties:
SYSTem:COMMunicate:RDEVice:AKAL:DATE?
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Calibration Type / Source
Selects the calibration type for the selected physical ports. For an overview, refer to
Table 5-7.
The reflection calibration types can be used for any combination of physical ports:
reflection calibrations are performed for each selected port.
A transmission calibration type requires at least two physical ports. For the unidirec-
tional transmission calibration types ("Trans Norm", "One Path Two Ports"), the direc-
tion ("Source" port) must be specified in addition.
Note: Transmission normalization and "One Path Two Ports" calibration types require
two-port (Through) characterization data for the cal unit.For current calibration units,
however, these two-port characterizations are typically not part of the factory character-
ization. If a tooltip indicates missing two-port characterization data, simply perform a
new characterization of your cal unit. In the first dialog of the "Characterization" wizard,
select "Take All OSM and Through" to make sure that the necessary two-port data is
acquired. See also "Characterization Wizard" on page 488.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Cal Softtool