GUI Reference
R&S
®
ZNB/ZNBT
422User Manual 1173.9163.02 ─ 53
Common Port
This combo-box is only available if Use Reduced Number of Through is activated in the
"Calibration" tab of the "System Config" dialog.
It defines the port that must appear in each port assignment of an automatic calibra-
tion, and that will be used as the "center" of all Through measurements.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Select All/Deselect All
Selects/deselects all ports.
These buttons are only shown if more than 4 ports are available.
Next
Proceeds to Step 2: Connections. "Next" is unavailable (and a warning is displayed) if
the following happens:
The selected characterization data do not cover all the ports to be calibrated.
The selected characterization data do not contain all standards needed for the
selected calibration type.
Step 2: Connections
Defines the port assignments between the R&S ZNB/ZNBT and the calibration unit.
Background information
Refer to:
Chapter 5.5.5.4, "Multiple Port Assignments", on page 172 for details on multiple
port assignments
"Step 2: Connections" on page 455 for SMARTerCal.
Depending on the number of test ports, the user interface is slightly different:
Cal Softtool