GUI Reference
R&S
®
ZNB/ZNBT
483User Manual 1173.9163.02 ─ 53
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe:]CORRection:CKIT:CATalog?
Import Cal Kit... / Export Cal Kit...
The buttons below the "Connector Type" list are used to store cal kit data to a file and
to reload previously stored cal kit files. By default, calibration kit files are stored in the
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration directory;
see Chapter 5.5.2.4, "Cal Kit Files", on page 165.
Remote command:
MMEMory:LOAD:CKIT
MMEMory:STORe:CKIT
Kit Standards Dialog
The "Kit Standards" dialog shows the calibration standards in a selected calibration kit.
It is also used to modify the contents of a user-defined kit.
Related information
Refer to the following sections:
●
Chapter 5.5.2.4, "Cal Kit Files", on page 165
●
Chapter 6.11.2.2, "Calibration Kits Dialog", on page 481
●
Chapter 5.5.2.1, "Calibration Standard Types", on page 161
Access: Calibration Kits Dialog > "Standards..."
One port and two port standards are listed in two separate tables. Most of the buttons
on the right side are available only if the "Kit Standards" dialog was opened for a user-
defined calibration kit.
Cal Softtool