EasyManua.ls Logo

Rohde & Schwarz R&S ZNB - Page 541

Rohde & Schwarz R&S ZNB
1451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI Reference
R&S
®
ZNB/ZNBT
541User Manual 1173.9163.02 ─ 53
"Add CW Mode"
Activates a new channel for the spectrum measurement. The "CW Frequency" plus
half the "Tone Distance" defines the center of the diagram.
"Max IM Order"
Defines the width of the spectrum measurement.
Remote command:
[SENSe<Ch>:]FREQuency[:CW]
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum:MORDer
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:STATe]
6.12.4.2 Intermodulation Presetting Wizard
The "Intermodulation Presetting" wizard guides the user through the setup of an inter-
modulation measurement.
Access: CHANNEL – [CHANNEL CONFIG] > "Intermod" > "Intermod Wizard..."
The wizard proceeds in 4 steps:
1. Configure intermodulation measurement. Refer to Chapter 6.12.4, "Intermod. Tab",
on page 539.
Channel Config Softtool

Table of Contents