GUI Reference
R&S
®
ZNB/ZNBT
599User Manual 1173.9163.02 ─ 53
●
If the "Fixture Simulator" is disabled for the related channel (see "Fixture Simulator"
on page 506), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
●
For a R&S ZNB without second internal source, the
Overview
See "Overview" on page 572.
Type
Switches between "Deembedding" and "Embedding" network definition.
Network
The graphical list contains all available 1-port networks:
The
symbol selects "no network" and disables ground loop de-/embedding.
The "1-Port Data" network is defined by imported S-parameter data; see Select File...
The remaining networks (Shunt L and Shunt C) are defined by lumped elements
whose parameters are displayed below the graphical list.
Tip:
●
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol.
●
Use the Conductance in Embedding Networks switch to change from resistances
in "Capacitor in parallel with resistor" model to conductances and vice versa.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:C
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:L
Offset Embed Softtool