EasyManua.ls Logo

Rohde & Schwarz R&S ZNB - Page 842

Rohde & Schwarz R&S ZNB
1451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
842User Manual 1173.9163.02 ─ 53
Example: Suppose that the active recall set contains an active trace no. 1.
CALC:MARK:REF ON; :CALC:MARK ON
Create the reference marker and marker 1 and assign them to
trace no. 1. The default position of both markers is the center of
the sweep range.
Manual operation: See "On" on page 363
CALCulate<Chn>:MARKer<Mk>:REFerence:FORMat <OutFormat>
Sets/queries the output format for the (complex) value of the reference marker.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> This numeric suffix is ignored and may be set to any value.
Parameters:
<OutFormat> DEFault | MLINear | MLOGarithmic | MDB | PHASe | POLar |
COMPlex | GDELay | REAL | IMAGinary | SWR | LINPhase |
MLPHase | LOGPhase | MDPHase | IMPedance | ADMittance |
MIMPedance
DEFault means the reference marker is formatted according
the related trace's default marker format (see
CALCulate<Chn>:MARKer:DEFault:FORMat).
For the other marker formats see CALCulate<Chn>:MARKer:
DEFault:FORMat and the description in "Marker Format"
on page 98.
*RST: DEFault
Manual operation: See "Marker Format" on page 366
CALCulate<Chn>:MARKer<Mk>:REFerence:MODE <Mode>
Sets the reference marker to continuous or discrete mode. The marker doesn't have to
be created before (CALCulate<Chn>:MARKer<Mk>[:STATe] ON), the mode can be
assigned in advance.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> This numeric suffix is ignored and may be set to any value.
Parameters:
<Mode> CONTinuous | DISCrete
CONTinuous - marker can be positioned on any point of the
trace, and its response values are obtained by interpolation.
DISCrete - marker can be set to discrete sweep points only.
*RST: CONT
SCPI Command Reference

Table of Contents

Related product manuals