Digital Standard PHS SMIQ
1125.5555.03 E-92.128
(SELECT SLOT) R
Display of data contents in the 4-bit data field "Ramp".
SS
Display of data contents in the 2-bit data field
"Start Symbol".
PR
Display of "Preamble".
UW
Input value for the "Unique Word" in hexadecimal form.
The length of the hexadecimal value depends on the slot
type (16 | 32 bit).
IEC/IEEE-bus
:SOUR:PHS:SLOT2:UWOR #H3D4C
CI
Display of the "Channel Identifier" data field in hexa-
decimal form.
SA
Selection of data source for SACCH. This data field is
only displayed for burst types TCH and VOX.
PN.. PRBS data according to CCITT V52 or Rec.
0.151 with periods between 2
9
-1 and 2
23
-1.
IEEE
:SOUR:PHS:SLOT3:SACCH PN9
DLIST Data from a programmable data list.
IEEE
:SOUR:PHS:SLOT3:SACCH DLIS
SERDATA Data from data input SER DATA
IEEE
:SOUR:PHS:SLOT3:SACCH SDAT
TCH / VOX
Selection of data source for TCH data field. This data
field is only displayed for burst types TCH and VOX.
PN.. PRBS data according to CCITT V52 or Rec.
0.151 with periods between 2
9
-1 and 2
23
-1.
IEEE
:SOUR:PHS:SLOT3:TCH PN9
DLIST Data from a programmable data list.
IEEE
:SOUR:PHS:SLOT3:TCH DLIS
SERDATA Data from data input SER DATA
IEEE
:SOUR:PHS:SLOT3:TCH SDAT
CS-ID
Input value for the "Cell Station ID Code" field in hexa-
decimal form. This data field is only displayed for the
burst type SYNC.
IEEE-bus
:SOUR:PHS:SLOT2:CSID #H20200020001