SMIQ Digital Standard IS-95 CDMA
1125.5555.03 E-92.139
REV LINK FILTER...
Opens a window for selecting the baseband filter for
the reverse link. A selection between the FIR filters
defined in IS-95, the Nyquist filters COS and
SQRCOS or a user-defined filter USER (cf. Section
Digital Modulation) is possible. A FIR filter is preset
according to IS-95.
IEEE command :SOUR:IS95:FILT:RTYP IS95
ROLL OFF FACTOR
Setting value of roll-off factor.
The setting is only possible provided that a COS or
SQRCOS filter is selected under FILTER.
IEEE-bus command :SOUR:IS95:FILT:PAR 0.22
FILTER MODE
Selection of filter mode.
LOW_ACP Filter for minimum Adjacent C
hannel
Power.
IEEE SOUR:IS95:FILT:MODE LACP
LOW_EVM Filter for minimum vector error.
IEEE SOUR:IS95:FILT:MODE LEVM