Digital Standard IS-95 CDMA SMIQ
1125.5555.03 E-92.140
TRIGGER MODE...
Opens a window for selecting the CDMA sequence.
AUTO The calculated CDMA chip sequence is cyclically
repeated.
IEC/IEEE-bus command :SOUR:IS95:SEQ AUTO
RETRIG The CDMA chip sequence is continuously repeated. A
trigger event causes a restart from frame 1.
IEC/IEEE-bus command :SOUR:IS95:SEQ RETR
ARMED_AUTO The CDMA chip sequence cannot be started from frame
1 until a trigger event has occurred. The unit then
automatically switches over to the AUTO mode and can
no longer be triggered.
IEC/IEEE-bus command :SOUR:IS95:SEQ AAUT
ARMED_RETRIG The CDMA chip sequence cannot be started from frame
1 until a trigger event has occurred. each new trigger
event causes a restart.
IEC/IEEE-bus command :SOUR:IS95:SEQ ARET
EXECUTE TRIGGER
Executes a trigger event to start the CDMA chip sequence.
IEC/IEEE-bus command :TRIG:DM:IMM
TRIGGER...
Opens a window for selecting the trigger source, for configuring the trigger
output signals and for setting the time delay of an external trigger signal.
FREQ
100. 000 000 0
MHz
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
IS95
PHS
IS-95
STATE
MODE
SET DEFAULT
MODULATION...
TRIGGER MODE
EXECUTE TRIG
TRIGGER....
CLOCK...
TOTAL POWER
ADJUST TO TO
TRIGGER SOURCE
INT EXT
EXT TRIGGER DELAY 0 Symb
EXT RETRIGGER INHIBIT 0 Symb
TRIGGER OUT1 EVEN SECOND
TRIGGER OUT2 SHORT SEQ ROLLOVER
TRIGGER OUT1 POL POS NEG
TRIGGER OUT2 POL POS NEG
TRIGGER OUT1 DELAY 0
TRIGGER OUT2 DELAY 0
LEVEL
-30.0
dBm
PEP
- 13.9.
dBm
Fig. 2-83 Menu DIGITAL STD - IS-95 - TRIGGER..., equipped with options modulation coder
SMIQB20, data generator SMIQB11 and SMIQB42