EasyManua.ls Logo

Sabio 2010D - Predefined Operator Stepped Manual Calibration; Manual Sequence Selection Menu

Sabio 2010D
182 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Page 5-
3
Predefined Operator Stepped Manual Calibration
To perform a manual calibration using predefined or programmed
sequences, perform the following steps:
[1] From the Main Menu screen, select Sequences and press the
Enter button. A pop-up menu will appear.
[2] Select Run, and press the Enter button. Another pop-up
menu will appear that allows the choice of which type of
manual calibration you wish to use.
[3] Select Operator Stepped from the menu and press the Enter
button. A list of sequences that have been predefined or
programmed will appear (see Figure 5-1).
Figure 5-1 Manual Sequence Selection Menu
[4] Select the sequence containing the point you want to run.
The "Select a Point to Run" screen shown in Figure 5-2 will
appear, displaying the Sequence Name and Type, the Primary
Gas Name and concentrations for each point in the sequence
and a box entitled "Run Point Number", which will allow the
user to select a calibration point to run.

Table of Contents

Other manuals for Sabio 2010D