Model 2010D Calibration Setup Page 6-
20
[23] Repeat steps 21 and 22 until all the Instrument Solenoids
and Point Status Outputs have been set up for each
calibration point or until all 10 points on the page have
been entered. If there more than 10 points in this
sequence, press F5 after you have entered the first 10
points. This takes you to page 4, which allows you to
program an additional 10 calibration points.
[24] When all points have been defined, the sequence information
can be reviewed by pressing F4 and F5 to page up and page
down. Information can be edited by positioning the cursor
using the F2 and F3 key to move backward or forward between
fields and then entering the new information in the
highlighted field.
When finished, press End Sequence from any page. A message
will pop up asking if you want to save the sequence.
Responding "yes" will save the sequence information you
have entered into non-volatile memory. Answering "no" will
exit, discarding any information that has been entered.
Pressing End Sequence again will return you to the New
Sequence screen without saving or discarding information.