5.4 Pattern copy/delete & Segment insert/delete
Program pattern copy
Set object pattern number on D1000
Set target pattern number on D1001
Set PT COPY TRIGGER('4') on D1003
Read D1004 data and then wait until DONE('1')
Program pattern delete
Set delete pattern number on D1000
Set PT DEL TRIGGER('5') on D1003
Read D1004 data and then wait until DONE('1')
Insert segment
Set program pattern number on D1000
Set insert pattern number on D1002
Set Temp, Humi TSP and segment time on D1010~D1012
Set SEG INSERT TRIGGER('6') on D1003
Read D1004 data and then wait until DONE('1')
Delete segment
Set program pattern number on D1000
Set delete segment number on D1002
Set SEG DEL TRIGGER('7') D1003
Read D1004 data and then wait until DONE('1')