MULTI-PORT OPTICAL POWER METER MPM-210H
31 / 59
0 Internal trigger
1 External trigger
If it is internal trigger, it generates trigger every 10ms.
Default
1
Example
TRIG 1
2. TRIG?
Description
If measuring mode is ‘CONSTx’, it checks for measuring
trigger.
Syntax
TRIG?
Response
<value>
0 Internal trigger
1 External trigger
Example
TRIG? 1
Response: 1
3. WMOD
Description
Set measurement mode.
The MPM-215 supports only “CONST2” and “SWEEP2”.
Syntax
WMOD<wsp><value>
Parameters
<value>
“CONST1” => Constant Wavelength, No Auto Gain, SME
mode
“SWEEP1” => Sweep Wavelength, No Auto Gain, SME
mode
“CONST2” => Constant Wavelength, Auto Gain, SME
mode
“SWEEP2” => Sweep Wavelength, Auto Gain, SME mode
“FREERUN” => Constant Wavelength, No Auto Gain, First
Hardware Trigger Start (CEM mode)
Refer to the 5.6.1 WMOD.
Default
CONST1
Example
WMOD CONST1
4. WMOD?