MPM-210H OPERATION MANUAL
32 / 59
Description
Check measurement mode.
The MPM-215 supports only “CONST2” and “SWEEP2”.
Syntax
WMOD?
Response
“CONST1” => Constant Wavelength, No Auto Gain, SME
mode
“SWEEP1” => Sweep Wavelength, No Auto Gain, SME
mode
“CONST2” => Constant Wavelength, Auto Gain, SME
mode
“SWEEP2” => Sweep Wavelength, Auto Gain, SME mode
“FREERUN” => Constant Wavelength, No Auto Gain, First
Hardware Trigger Start (CEM mode)
Refer to the 5.6.1 WMOD.
Example
WMOD?
Response: CONST1
5. WAV
Description
Set up for using wavelength of Constant wavelength
Measurement Mode (CONST1, CONST2, READ).
Syntax
WAV<wsp><value>
Parameters
<value>
1250.000 ~ 1630.000 Real number
Unit: nm
Default
1550
Example
WAV 1550
6. WAV?
Description
Check using wavelength of Constant wavelength
Measurement Mode (CONST1, CONST2, READ).
Syntax
WAV?
Response
<value>
1250.000 ~ 1630.000 Real number
Unit: nm
Example
WAV?
Response: 1550