MULTI-PORT OPTICAL POWER METER MPM-210H
33 / 59
7. DWAV
Description
Set up for using wavelength of Constant wavelength
Measurement Mode (CONST1, CONST2, READ) for each
channel.
Syntax
DWAV<wsp><value1>,<value2>,<value3>
Parameters
<value1>: Module 0, 1, 2, 3, 4, 5
When setting the <value 1> to 5, all channels and
module is set to the same wavelength.
<value2>: Channel 1, 2, 3, 4
<value3>: 1250.000 ~ 1630.000 Real number
Unit: nm
Default
1550
Example
DWAV 0,1,1550
8. DWAV?
Description
Check using wavelength of Constant wavelength
Measurement Mode (CONST1, CONST2, READ) for each
channel.
Syntax
DWAV? <wsp><value1>,<value2>
Parameters
<value1>: Module 0,1, 2, 3, 4
<value2>: Channel 1, 2, 3, 4
Response
<value>
1250.000 ~ 1630.000 Real number
Unit: nm
Example
DWAV? 0,1
Response: 1550
9. WSET
Description
Set up for using wavelength of Sweep Wavelength
Measurement Mode (SWEEP1, SWEEP2).
The data points is calculated from the condition that is set
by the “WSET” command and it is possible to check the
data points by the “LOGN?” command. Please refer to
“5.6.3. Measurement data point setting (LOGN)”.