MPM-210H OPERATION MANUAL
34 / 59
Syntax
WSET<wsp><value1>,<value2>,<value3>
Parameters
<value1>: Start wavelength 1250 ~ 1630
<value2>: Stop wavelength 1250 ~ 1630
<value3>: Step wavelength 0.001 ~ 10
Unit: nm
Note) Stop wavelength = Start wavelength + step
wavelength
Default
value1: 1250, value2: 1630, value3: 0.005
Example
WSET 1520,1570,0.05
10. WSET?
Description
Check using wavelength of Sweep Wavelength
Measurement Mode (SWEEP1, SWEEP2).
Syntax
WSET?
Response
<value1>,<value2>,<value3>
<value1>: Start wavelength 1250 ~ 1630
<value2>: Stop wavelength 1250 ~ 1630
<value3>: Step wavelength 0.001 ~ 10
Unit: nm
Example
WSET?
Response: 1520,1570,0.05
11. SPE
Description
Set up Wavelength Sweep Speed (Sweep Wavelength
Measurement Mode (SWEEP1, SWEEP2)). The averaging
time is automatically set by the measurement speed and
measurement step of WSET.
Syntax
SPE<wsp><value>
Parameters
<value>: Sweep speed 0.001 ~ 200
Unit: nm/sec
Default
value: 1
Example
SPE 1
12. SPE?
Description
Check Wavelength Sweep Speed (Sweep Wavelength