User manual SCAIME: NU-soft-eNod4B-E-0716_236704-C.doc
13.2 Settings description
13.2.1 Logical inputs assignment
The following table describes the possible assignments.
bits
meaning
note
b
3
b
2
b
1
b
0
input 1&3 assignment
0000
none
the input has no function
0001
tare
equivalent to the functions described in §
functional commands
0010
zero
0011
cancel tare
0100
transmit measurement*/send TPDO2**
data is transmitted on the bus at every
rising or falling edge (depending on the
chosen logical) Transmitter mode only
0101
measurement window*
Data is transmitted on the bus while the
input is maintained at the right level
(depending on the chosen logical).
Transmission rate is fixed by the ‘sampling
rate’ setting
0110
Dynamic zero
Same like equivalent functional command
describe in § functional commands
0111
Start / Stop
Belt mode
1000
Belt running detection
1001
Clear totalization and errors counter
Same like equivalent functional command
describe in § functional commands
1010
Sensor input control
Same like equivalent functional command
describe in § functional commands
1011
Belt fault
In this case, the input is linked to fault
sensor. This function triggers alarm.
b
4
input 1&3 logical
0
negative logic
defines the edge (or level) that triggers
input function
1
positive logic
b
6
b
5
measurement to be transmitted
00
gross
only for SCMBus/fast SCMBus protocols, no
effect otherwise
01
net
10
factory calibrated measurement
bits
meaning
note
b
11
b
10
b
9
b
8
(or b
3
b
2
b
1
b
0
in CANopen®)
input 2&4 assignment
0000
none
the input has no function